R / C. No.: Autoclave/CH-2/RC-C2010000/0306/0010/026/CCOAD Dated : 22.03.2005.
To
M/s. Yorco Scientific Udyog Pvt. Ltd.,
No.13, Loni Road, Site 2nd Industrial Area, BY SPEED POST
Mohan Nagar,
Ghaziabad-207 007 (U.P.)
|
Subject : |
Rate Contract for supply of Autoclave Steam Sterilisers valid From 01.04.2005 to 31.03.2006.
|
|
Reference : |
1. This office tender enquiry No. Autoclave/CH-2/RC-C2010000/0306/ 0010 opened on 29.12.2004. 2. Your Quotation No. YU/Q-2094/CS dated 22.12.2004 and subsequent letter No.YU/Q-2087/US dated 09.02.2005 and YSU/Q-2087/2004-05/SKG dated 12.03.2005. |
|
Dear Sirs, |
|
You are hereby informed that your above referred tender read with subsequent letters mentioned above for the stores specified in the Schedules annexed has been accepted. This Rate Contract will be governed by the terms and conditions brought out in the form no. DGS&D-1001. The Rate Contract and the Schedules annexed hereto shall be the sole repository of this RATE CONTRACT / TRANSACTION. |
|
Schedules annexed: |
|
|
Yours faithfully, |
|
( C. ARUMUGAM ) Assistant Director (Supplies ) For and on behalf of the Purchaser(s) named in the form DGS&D - 1001 |
Copy to:
|
SCHEDULE ‘A’ |
|||
|
|
|||
|
1 |
Rate Contract No. Autoclave/CH-2/RC-C2010000/0306/0010/026/CCOAD Dated : 22.03.2005 for the supply of Autoclave Steam Sterilisers. |
||
|
2 |
Advance Rate Contract No. Nil dated Nil |
||
|
3 |
a) |
Name and full address of the firm: |
M/s. Yorco Scientific Udyog Pvt. Ltd., No.13, Loni Road, Site 2nd Industrial Area, Mohan Nagar, Ghaziabad-207 007 (U.P.) |
|
Telephone No.: |
91-0120-2940533, 2940535 |
||
|
Fax No.: |
91-0120-2940739 |
||
|
E-mail address: |
steri@tatanova.com |
||
|
b) |
Name and address of manufacturer: |
Same as above |
|
|
Telephone No.: |
Same as above |
||
|
Fax No.: |
Same as above |
||
|
4 |
Validity of Rate Contract : |
From 01.04.2005 to 31.03.2006 |
|
|
5 |
Description of item, specification, unit rate, etc. |
||
|
Schedule No. (1) |
Description of item (2) |
Unit (3) |
Rate per unit (4) |
||||||||||||||
|
1 |
Horizontal Cylindrical high speed steam
sterilizer pressure type BIS marked,
High speed horizontal cylinderical steam sterilizers pressure type of size
400 mm dia x 600 mm length designed to operate on steam from stem generator as
per IS: 3829 (Part 2)/1978 (reaffirmed 2001) with amendment Nos 1 to 4 and
shall be fitted with only silicon rubber door gasket (expansion type). The
steam generator shall be electrically heated by immersion heaters having mineral
filled sheathed heating elements conforming to IS:4159/2002 duly BIS marked
and complying with safety requirements as per IS:302-2-2001 (1992). The immersion
heaters shall be wired for operation on 3 phase 4 wire 400/440 V, 50 Hz, AC
supply and electric load for the unit shall be 18 KW. Air shall be
removed from the sterilization chamber after loading the sterilizer by one or
more pressure/vaccum pulsing for proper steam penetration either manually or
automatically through the automatic control system. Types of Steam
Sterilizers are as under:
|
No. |
1,35,700/- |
|
2 |
Horizontal
Cylindrical steam sterilizer pressure type for hospital and pharmaceutical
use BIS marked
pressure type steam sterilizers of horizontal cylinderical type. The steam
sterilizer shall conform to IS:3829 (Part 1)/1999 and shall be fitted with
only silicon rubber door gasket (expansion type). The sterilzer shall have
single door. The immersion heater used for electric heating shall be BIS
marked and shall conform to to IS:4159/2002 and complying with safety
requirements as per IS:302-2-201(1992).
The immersion Heater shall have mineral filled sheathed heating elements
and shall BIS marked conforming to:IS:4159/2002 and complying with safety
requirements as per IS:302-2-201(1992).The immersion heater shall be wired
for operation on 3 Phase, 4 Wire,400/440V A.C Supply. Air shall be
removed from the sterilization chamber after loading the sterilizer by one or
more pressure/vaccum pulsing for proper steam penetration. The sterilizer shall
be of the following size/type and corresponding electric loads.
|
No. No. No. No. No. No. No. No. No. |
1,45,117/- 1,69,117/- 1,30,117/- 1,64,308/- 1,54,308/- 1,24,308/- 1,45,117/- 1,69,117/- 1,30,117/- |
|
3 |
Horizontal Rectangular type/horizontal square
type steam sterilizers pressure type for hospital and pharmaceutical use BIS marked
pressure type steam sterilizers of horizontal Rectangular or Horizontal
Square shape. The steam sterilizer
shall conform to IS:3829 (Part 1)/1999 and shall be fitted with only silicon
rubber door gasket (expansion type). The sterilzer shall have single door. The immersion heater used for electric
heating shall have mineral filled sheathed heating elements and shall be BIS
marked conforming to IS:4159/2002 and complying with safety requirements as
per IS:302-2-201(1992). The immersion Heater shall be wired for operation on
3Phase, 4 Wire, 400/440V A.C Supply. Air shall be
removed from the sterilization chamber after loading the sterilizer by one or
more pressure/vaccum pulsing for proper steam penetration. The sterilizer shall be of the
following size/type and corresponding electric loads.
|
No. No. No. No. No. No. |
3,71,700/- 3,96,700/- 3,46,700/- 4,89,700/- 4,64,700/- 5,14,700/- |
|||||||||||||||||||||||||||||||||||||||||||||||||
|
4 |
Vertical Cylindrical type steam sterilizers
pressure type BIS marked
pressure type steam sterilizers of vertical cylinderical type with built in
steam generator. The steam generator shall be electrically heated using BIS
marked immersion heater having mineral filled sheathed heating elements conforming
to IS:4159/2002 and complying with safety requirements as per
IS:302-2-201(1992).The immersion Heater shall be wired for operation on
3Phase, 4 Wire,400/440V A.C Supply. The
steam sterilizer shall conform to IS:3829 (Part-3) 1985 reaffirmed 2001) with
Amdt. 1 & 2 and shall be fitted with only silicon rubber door gasket
(expansion type). |
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
|