Rate Contract No.:CH-2/RC-18019002/112003/AC/STEAM STERILIZERS /5/072/DCOAM

      Dated : 02.04.2004.

 

To

 

M/s. Reliance Instruments Corporation,

No.217, Sidco Industrial Estate,

North Phase, Ambattur,

CHENNAI-600 098.

 

BY REGD. POST A.D.

Subject:

Rate Contract for supply of AUTO CLAVE STEAM STERILIZERS  valid From 02.04.2004 TO 31.03.2005.

 

Reference:

(i) This office tender enquiry no.CH-2/RC-18019002/112003/ AC/Steam Sterilizers opened on 30.12.2003.

(ii) Your quotation No. RIC/QTN/241/2003-04 Dated 29.12.2004 and subsequent letters  No.RIC/GEN/527/2003-04 DATED 23.01.2004 and RIC/QTN/324/2003-04 dated 24.03.2004.

 

 

Dear Sirs,

 

                                You are hereby informed that your above referred tender read with subsequent letters mentioned above for the stores specified in the Schedules annexed has been accepted. This Rate Contract will be governed by the terms and conditions brought out in the form no. DGS&D-1001. The Rate Contract and the Schedules annexed hereto shall be the sole repository of this Rate Contract/Transaction.

 

Schedules annexed:

 

  1. Schedule ‘A’ - Description of Stores, prices, duties/taxes
  2. Schedule ‘B’  - Information to DDOs about RateContract.
  3. Schedule ‘C’ -  List of Parallel Rate Contract Holders.

 

Yours faithfully,

 

 

 

(G.M.RAO)

ASST DIRECTOR(SUPPLIES)

For and on behalf of the Purchaser(s) named in the form DGS&D – 1001

 

 

--  2  --

 

 

Copy to:

 

1.       The Deputy Controller of Accounts, Department of Commerce (Supply Division), Chennai.           

       (Through Authentication).

2.       Directorate General of Supplies & Disposals, New Delhi.

3.       Deputy Director General of Supplies & Disposals, Kolkata.

4.       Deputy Director General of Supplies & Disposals, Mumbai.

5.       Deputy Director General (QA), Chennai.

6.       Deputy Director General (QA), Kolkata.

7.       Deputy Director General (QA), Mumbai.

8.       Deputy Director General (QA), DGS&D, New Delhi.

9.       Inspection Authority : A.D.G.(Q.A.), DGS&D., New Delhi.

10.    Quality Assurance Officer : Director (Q.A.),D.G.S.& D., Shastri Bhavan, CHENNAI-6.

11.    Concurrent Audit

12.    MIS Cell

13.    Ledger Clerk

14.    Computer Section.

15.    All Direct Demanding Officers as per mailing list maintained.

16.    Deputy Director of Supplies.

17.    CDN section.

18.    Master Folder.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

--   3   --

 

SCHEDULE ‘A’

 

 

 

 

 

1

Rate Contract No.:CH-2/RC-18019002/112003/AC/STEAM STERILIZERS /5/ 072/DCOAM           

Dated 02.04.2004 for supply of AC STEAM STERILIZERS.

 

 

2

Advance Rate Contract No. NIL Dated Nil.

 

 

3

a)

Name and full address of the firm:

M/s. Reliance Instruments Corporation,

No.217, Sidco Industrial Estate,

North Phase, Ambattur,

CHENNAI-600 098.

 

 

 

 

 

 

 

 

Telephone No.:

044-26247124

 

 

Telex No.:

--

 

 

Fax No.:

91-044-26244393

 

 

E-mail address:

reico@satyam.net.in

 

 

b)

Name and address of manufacturer:

Same as above.

 

 

 

 

 

Telephone No.:

Same as above

 

 

Telex No.:

--

 

 

Fax No.:

Same as above.

 

 

4

Validity of Rate Contract:

From 02.04.2004 to 31.03.2005.

 

 

5

Description of item, specification, unit rate, etc.

 

 

Sl.

No.

Item No.

Description of item & Specification

Unit

Rate per unit

 

(1)

(2)

(3)

(4)

(5)

 

 

 

1.

 

I (a)

 

Horizontal Cylindrical High Speed Steam Sterilizers

Pressure Type :

 

 High Speed Pressure type steam sterilizers of horizontal cylindrical type of size         400 mm dia x 600mm length  designed to operate on steam from steam generator as per IS:3829 (Part 2 )/1978 (Reaffirmed 2001) with Amdt. Nos. 1 to 4 with provision of silicon door gasket (expansion type) for the sterilizers.  The steam generators shall be electrically heated by immersion heaters having mineral filled sheathed heating elements conforming to IS:4159/2002 duly ISI marked and complying with safety requirements as per IS:302-2-201(1992).  The immersion heaters shall  be wired for operation on 3 Phase 4 wire, 400/440, 50Hz,  A/C                 supply and the electric load for the unit shall be 18 KW.

                                                                                               

The designed working pressure is 220 KN/m2 (2.2 kg/cm2approx.) but set to             operate at 210(+)0(-) 15KN/m2 (2.1 +0-0.15Kg/cm2 approx.) and the temperature   of  saturated steam inside the chamber is 134’C.

 

 

No.

 

1,14,000/-

 

 

 

--  4  --

 

MATERIAL: THE MATERIAL OF CONSTRUCTION IS AS FOLLOWS:-

 

1.

Inner chamber

SS316L

2.

Steam Jacket

SS304

3.

Back Plate

SS316L

4.

Steam Generator

SS304

5.

Insulation cover & piping

SS304

6.

Door

SS Lined

 

NOTE:  YOU SHALL BE FULLY RESPONSIBLE FOR INSTALLATION, COMMISSIONING AND SATISFACTORY DRAWL FOR AT LEAST 15 CYCLES OVER A PERIOD OF 3 OR 4 DAYS AT CONSIGNEES PREMISES.

 

 

Sl.

No.

Item No.

Description of item & Specification

Unit

Rate per unit

(1)

(2)

(3)

(4)

(5)

1.

II(a)

Horizontal Cylindrical Steam Sterilizers Pressure Type              for Hospital and Pharmaceutical use :

 

Pressure type steam sterilizers of horizontal cylindrical type designed for the working pressure of 126 KN/m2  (1.26 Kg/cm2 approx.) but set to operate at 105      (+) 15 KN/m2 (1.05 + 0.15 Kg/cm2 approx.) and the temperature inside is about 121’C with built in steam generator which shall be electrically heated by immersion  heaters having mineral filled sheathed heating elements conforming to IS:4159/2002 duly ISI marked and complying with safety requirements as per IS: 302-2-201 (1992).  The immersion heaters shall be wired for operation in 3 Phase, 4 Wire 400/440 Volts, 50Hz, A/C supply.  The horizontal cylindrical steam sterilizer shall conform to IS:3829(Part 1)/ 1999 with provision of silicon door gasket (expansion type) for the sterilizers and of  the following sizes and with corresponding electric loads :

 

 

Sl.No.

Dia.(mm)

Length(mm)

Electric Load(KW)

i.

400

1100

9

ii.

500

900

9

iii.

500

1100

9

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No.

1,23,000/-

No.

1,23,000/-

No.

1,30,992/-

 

 

 

 

 

 

 

 

 

 

2.

III (a)              

Horizontal Rectangular type / Horizontal Square type steam sterilizers pressure type for hospital & pharmaceutical use :

 

Same as II(a) above but of Horizontal Square or Horizontal Rectangular type in    the following sizes and with corresponding electrical load and conforming to IS:3829 (Part 1) / 1999.

Sl.

No.

Width

(mm)

Height (mm)

Length (mm)

Electric load(KW)

 

i.

600

600

1200

18

 

ii.

600

900

1500

36