GOVERNMENT
OF MINISRY
OF COMMERCE & INDUSTRY DEPARTMENT
OF COMMERCE (SUPPLY WING) OFFICE
OF THE DIRECTOR OF SUPPLIES & DISPOSALS SHASTRI
BHAVAN, 26, TELEPHONE:
28273721/28275154 FAX: 044-28280152 www.dgsnd.tn.nic.in dsnd@tn.nic.in | |
|
Rate
Contract No.:CH-2/RC-18019002/112003/AC/STEAM STERILIZERS /1/
048/CCOAD |
Dated :
05.03.2004. |
To
|
M/s. OSHO Home
Appliances, 2H, EPIP, Jharmajri
(BADDI), P.O. BAROTIWALA
DISTRICT, PIN:174
103. |
BY REGD. POST
A.D. | |
|
Subject: |
Rate Contract for
supply of AUTO CLAVE STEAM STERILIZERS valid From 20.03.2004
31.03.2005. | |
|
Reference: |
(i) This office
tender enquiry no.CH-2/RC-18019002/112003/ AC/Steam Sterilizers opened on
30.12.2003. (ii) Your quotation
No. OSHO/DGS&D/CH/RC/STER/03 dt.26.12.2003 and subsequent letters
OSHO/DGS&D/CH/RC/STER/2004 dated 30.01.2004 and your fax
No.OSHO/DGS&D/CH/RC/STER/2004 Dated 24.02.2004. | |
Dear
Sirs,
You are hereby informed that your above referred tender read with
subsequent letters mentioned above for the stores specified in the Schedules
annexed has been accepted. This Rate Contract will be governed by the terms and
conditions brought out in the form no. DGS&D-1001. The Rate Contract and the
Schedules annexed hereto shall be the sole repository of this Rate
Contract/Transaction.
Schedules
annexed:
Yours
faithfully,
(G.M.RAO)
ASST
DIRECTOR(SUPPLIES)
For and on behalf of the
Purchaser(s) named in the form DGS&D – 1001
-- 2
--
Copy
to:
1.
The Chief Controller of
Accounts, Department of Commerce (Supply Division),
(Through
Authentication).
2.
Directorate General of
Supplies & Disposals,
3.
Deputy Director General of
Supplies & Disposals, Kolkata.
4.
Deputy Director General of
Supplies & Disposals, Mumbai.
5.
Deputy Director General
(QA), Chennai.
6.
Deputy Director General
(QA), Kolkata.
7.
Deputy Director General
(QA), Mumbai.
8.
Deputy Director General
(QA), DGS&D,
9.
Inspection Authority :
A.D.G.(Q.A.), DGS&D.,
10. Quality Assurance Officer :
Director (Q.A.),D.G.S.& D., Mohali.
11. Concurrent
Audit
12. MIS
Cell
13. Ledger
Clerk
14. Computer
Section.
15. All Direct Demanding
Officers as per mailing list maintained.
16. Deputy Director of
Supplies.
17. CDN
section.
18. Master
Folder.
-- 3 --
|
SCHEDULE
‘A’ |
|
| |||||||
|
|
|
| |||||||
|
1 |
Rate Contract
No.:CH-2/RC-18019002/112003/AC/STEAM STERILIZERS /1/ 048/CCOAD
Dated 05.03.2004 for
supply of AC STEAM STERILIZERS. |
|
| ||||||
|
2 |
Advance Rate Contract
No. NIL Dated Nil. |
|
| ||||||
|
3 |
a) |
Name and full address
of the firm: |
M/s. OSHO Home
Appliances, 2H, EPIP, Jharmajri
(BADDI), P.O. BAROTIWALA
DISTRICT, PIN:174
103. |
|
| ||||
|
|
|
||||||||
|
|
|
||||||||
|
|
Telephone
No.: |
01792-203741,
200672-73, 09816044234, 098162-44234 |
|
| |||||
|
Telex
No.: |
-- |
|
| ||||||
|
Fax
No.: |
-- |
|
| ||||||
|
E-mail
address: |
Oshohome
app@indiatimes.com |
|
| ||||||
|
b) |
Name and address of
manufacturer: |
Same as
above. |
|
||||||
|
|
|
||||||||
|
|
Telephone
No.: |
Same as above
|
|
| |||||
|
Telex
No.: |
-- |
|
| ||||||
|
Fax
No.: |
-- |
|
| ||||||
|
4 |
Validity of Rate
Contract: |
From 20.03.2004 to
31.03.2005. |
|
| |||||
|
5 |
Description of item,
specification, unit rate, etc. |
|
| ||||||
|
Sl. No. |
Item
No. |
Description of item
& Specification |
Unit |
Rate per
unit |
| ||||
|
(1) |
(2) |
(3) |
(4) |
(5) |
| ||||
|
| |||||||||
|
1. |
I
(a) |
Horizontal
Cylindrical High Speed Steam Sterilizers Pressure Type :
High Speed Pressure type steam
sterilizers of horizontal cylindrical type of size
400 mm dia x 600mm length
designed to operate on steam from steam generator as per IS:3829
(Part 2 )/1978 (Reaffirmed 2001) with Amdt. Nos. 1 to 4 with provision of
silicon door gasket (expansion type) for the sterilizers. The steam generators shall be
electrically heated by immersion heaters having mineral filled sheathed
heating elements conforming to IS:4159/2002 duly ISI marked and complying
with safety requirements as per IS:302-2-201(1992). The immersion heaters shall be wired for operation on 3 Phase
4 wire, 400/440, 50Hz, A/C
supply and the electric load for the unit shall be 18
KW.
The designed working
pressure is 220 KN/m2 (2.2 kg/cm2approx.) but set to
operate at 210(+)0(-) 15KN/m2 (2.1 +0-0.15Kg/cm2 approx.) and the
temperature of saturated steam inside the chamber
is 134’C. |
No. |
90,500/- |
-- 4
--
|
2. |
V. |
Sterilizer, Portable,
vertical, Pressure type (Electrical Operated) Portable pressure
type vertical sterilizer shall have a chamber of diameter 350 mm and
overall height 300 to 325 mm, heated externally and operating under
saturated steam at a working pressure of 1.86 kgf/Cm2. It shall conform to IS:8462/1977
(Reaffirmed 2001) with Amdt.No.1 & 2 except that there shall be no
provision for heating by gas/kerosene or Charcoal as per clause 3 of IS
Specification. The Sterilizer
shall operate only electrically on single phase 230-250 volts 50Hz AC and
electric load shall be 2 KW. It shall have a
readily replaceable immersion type heaters having mineral filled sheathed
heating elements conforming to IS:4159/2002 duly ISI Marked and complying
with safety requirements as per IS:302-2-201 (1992). The SS Trays,
Shelves, baskets or shallow sterilizers as per Clause 5.7 of IS:8462/1997
shall be of adequate strength for sterilization and shall be quoted as
optional extra by the tenderer. Size: 350 mm x 300 to
325
depth |
Each |
4,450/- | |||||||||||||||
|
3. |
VI |
Sterilizer,
Instruments, table Model (electrically Operated) The Instrument
Sterilizer for disinfecting different types of instruments used in clinics
and hospitals shall conform to IS:5022/1989 (Amdt.No.1 & 2) Reaffirmed
in 1999 and in the following sizes :-
The Sterilizer shall
operate Electrically only on single phase 230-250 Volts 50Hz AC and there
shall be no provisions for Gas/Kerosene operation as per Clause 8.3/8.4 of
IS:5022/89. The Sterilizer
shall have replaceable immersion type heaters having mineral filled
sheathed heating elements conforming IS:4159/2002 duly ISI marked and
complying with safety requirements as per IS:302-2-201
(1992). |
|
| |||||||||||||||
|
Each |
950/- | ||||||||||||||||||
|
Each |
1850/- | ||||||||||||||||||
|
|
|
-- 5
--
|
(i) |
Materials used in the
Fabrication of Sterilizers as per IS:3829 part
1/1999. (a) Chamber Stainless
Steel X04Cr19Ni9 or 04 Cr 17 Ni 12 MO2 and of 2 B finish in accordance
with Table 8 of IS:6911 | ||||
|
(ii) |
Steriliser,
Instruments, Table Model IS:5022/1989 Material: 3.1. All the components other than the
drain cock, the element and the thermostat shall be made from stainless
steel shall conform to Designation 04Cr18Ni10 of Table 1 and of 2B or 2D
finish as given in Table 8 of IS:6911/1972. NOTE: YOU SHALL BE FULL RESPONSIBLE FOR
INSTALLATION, COMMISSIONING AND SATISFACTORY DRAWL FOR AT LEAST 15 CYCLES
OVER A PERIOD OF 3 OR 4 DAYS AT CONSIGNEES
PREMISES. | ||||
|
Sl. No. |
Item
No. |
Description of item
& Specification |
Unit |
Rate per
unit | |
|
(1) |
(2) |
(3) |
(4) |
(5) | |
|
1. |
II(a) |
Horizontal
Cylindrical Steam Sterilizers Pressure Type
for Hospital and Pharmaceutical use : Pressure type steam
sterilizers of horizontal cylindrical type designed for the working
pressure of 126 KN/m2 (1.26
Kg/cm2 approx.) but set to operate at 105 (+) 15 KN/m2
(1.05 + 0.15 Kg/cm2 approx.) and the temperature inside is about 121’C
with built in steam generator which shall be electrically heated by
immersion heaters having
mineral filled sheathed heating elements conforming to IS:4159/2002 duly
ISI marked and complying with safety requirements as per IS: 302-2-201
(1992). The immersion heaters
shall be wired for operation in 3 Phase, 4 Wire 400/440 Volts, 50Hz, A/C
supply. The horizontal
cylindrical steam sterilizer shall conform to IS:3829(Part 1)/ 1999 with
provision of silicon door gasket (expansion type) for the sterilizers and
of the following sizes and
with corresponding electric loads :
|
|
| ||||||||||||||||
|
No. |
77,500/- | |||||||||||||||||||
|
No. |
77,500/- | |||||||||||||||||||