Subject:

Rate Contract for supply of AUTO CLAVE STEAM STERILIZERS  valid From 20.03.2004 31.03.2005.

 

Reference:

(i) This office tender enquiry no.CH-2/RC-18019002/112003/ AC/Steam Sterilizers opened on 30.12.2003.

(ii) Your quotation No. OSHO/DGS&D/CH/RC/STER/03 dt.26.12.2003 and subsequent letters OSHO/DGS&D/CH/RC/STER/2004 dated 30.01.2004 and your fax No.OSHO/DGS&D/CH/RC/STER/2004 Dated 24.02.2004.

 

 

Dear Sirs,

 

                                You are hereby informed that your above referred tender read with subsequent letters mentioned above for the stores specified in the Schedules annexed has been accepted. This Rate Contract will be governed by the terms and conditions brought out in the form no. DGS&D-1001. The Rate Contract and the Schedules annexed hereto shall be the sole repository of this Rate Contract/Transaction.

 

Schedules annexed:

 

  1. Schedule ‘A’ - Description of Stores, prices, duties/taxes
  2. Schedule ‘B’  - Information to DDOs about RateContract.
  3. Schedule ‘C’ -  List of Parallel Rate Contract Holders.

 

Yours faithfully,

 

 

 

(G.M.RAO)

ASST DIRECTOR(SUPPLIES)

For and on behalf of the Purchaser(s) named in the form DGS&D – 1001

 

--  2  --

 

 

Copy to:

 

1.       The Chief Controller of Accounts, Department of Commerce (Supply Division), New Delhi.           

       (Through Authentication).

2.       Directorate General of Supplies & Disposals, New Delhi.

3.       Deputy Director General of Supplies & Disposals, Kolkata.

4.       Deputy Director General of Supplies & Disposals, Mumbai.

5.       Deputy Director General (QA), Chennai.

6.       Deputy Director General (QA), Kolkata.

7.       Deputy Director General (QA), Mumbai.

8.       Deputy Director General (QA), DGS&D, New Delhi.

9.       Inspection Authority : A.D.G.(Q.A.), DGS&D., New Delhi.

10.    Quality Assurance Officer : Director (Q.A.),D.G.S.& D., Mohali.

11.    Concurrent Audit

12.    MIS Cell

13.    Ledger Clerk

14.    Computer Section.

15.    All Direct Demanding Officers as per mailing list maintained.

16.    Deputy Director of Supplies.

17.    CDN section.

18.    Master Folder.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

--   3   --

 

SCHEDULE ‘A’

 

 

 

 

 

1

Rate Contract No.:CH-2/RC-18019002/112003/AC/STEAM STERILIZERS /1/ 048/CCOAD           

Dated 05.03.2004 for supply of AC STEAM STERILIZERS.

 

 

2

Advance Rate Contract No. NIL Dated Nil.

 

 

3

a)

Name and full address of the firm:

M/s. OSHO Home Appliances,

2H, EPIP, Jharmajri (BADDI),

P.O. BAROTIWALA DISTRICT,

PIN:174 103.

 

 

 

 

 

 

 

 

Telephone No.:

01792-203741, 200672-73, 09816044234, 098162-44234

 

 

Telex No.:

--

 

 

Fax No.:

--

 

 

E-mail address:

Oshohome app@indiatimes.com

 

 

b)

Name and address of manufacturer:

Same as above.

 

 

 

 

 

Telephone No.:

Same as above

 

 

Telex No.:

--

 

 

Fax No.:

--

 

 

4

Validity of Rate Contract:

From 20.03.2004 to 31.03.2005.

 

 

5

Description of item, specification, unit rate, etc.

 

 

Sl.

No.

Item No.

Description of item & Specification

Unit

Rate per unit

 

(1)

(2)

(3)

(4)

(5)

 

 

 

1.

 

I (a)

 

Horizontal Cylindrical High Speed Steam Sterilizers

Pressure Type :

 

 High Speed Pressure type steam sterilizers of horizontal cylindrical type of size         400 mm dia x 600mm length  designed to operate on steam from steam generator as per IS:3829 (Part 2 )/1978 (Reaffirmed 2001) with Amdt. Nos. 1 to 4 with provision of silicon door gasket (expansion type) for the sterilizers.  The steam generators shall be electrically heated by immersion heaters having mineral filled sheathed heating elements conforming to IS:4159/2002 duly ISI marked and complying with safety requirements as per IS:302-2-201(1992).  The immersion heaters shall  be wired for operation on 3 Phase 4 wire, 400/440, 50Hz,  A/C               supply and the electric load for the unit shall be 18 KW.

                                                                                               

The designed working pressure is 220 KN/m2 (2.2 kg/cm2approx.) but set to                 operate at 210(+)0(-) 15KN/m2 (2.1 +0-0.15Kg/cm2 approx.) and the temperature   of  saturated steam inside the chamber is 134’C.

 

 

No.

 

90,500/-

 

--  4  --

 

2.

 

V.

 

Sterilizer, Portable, vertical, Pressure type (Electrical Operated)

 

Portable pressure type vertical sterilizer shall have a chamber of diameter 350 mm and overall height 300 to 325 mm, heated externally and operating under saturated steam at a working pressure of 1.86 kgf/Cm2.  It shall conform to IS:8462/1977 (Reaffirmed 2001) with Amdt.No.1 & 2 except that there shall be no provision for heating by gas/kerosene or Charcoal as per clause 3 of IS Specification.  The Sterilizer shall operate only electrically on single phase 230-250 volts 50Hz AC and electric load shall be 2 KW.

 

It shall have a readily replaceable immersion type heaters having mineral filled sheathed heating elements conforming to IS:4159/2002 duly ISI Marked and complying with safety requirements as per IS:302-2-201 (1992).

 

The SS Trays, Shelves, baskets or shallow sterilizers as per Clause 5.7 of IS:8462/1997 shall be of adequate strength for sterilization and shall be quoted as optional extra by the tenderer.

 

Size:

350 mm x 300 to 325  depth

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Each

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4,450/-

 

3.

VI

Sterilizer, Instruments, table Model (electrically Operated)

The Instrument Sterilizer for disinfecting different types of instruments used in clinics and hospitals shall conform to IS:5022/1989 (Amdt.No.1 & 2) Reaffirmed in 1999 and in the following sizes :-

 

 

Length

Width

Depth

Elect.Load(KW)

 

a.

300 mm

150 mm

125 mm

1.00

c.

510 mm

200mm

150mm

2.00

 

The Sterilizer shall operate Electrically only on single phase 230-250 Volts 50Hz AC and there shall be no provisions for Gas/Kerosene operation as per Clause 8.3/8.4 of IS:5022/89.  The Sterilizer shall have replaceable immersion type heaters having mineral filled sheathed heating elements conforming IS:4159/2002 duly ISI marked and complying with safety requirements as per IS:302-2-201 (1992).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Each

950/-

Each

1850/-

 

 

 

 

 

 

--  5  --

 

 (i)

Materials used in the Fabrication of Sterilizers as per IS:3829 part 1/1999.

(a) Chamber Stainless Steel X04Cr19Ni9 or 04 Cr 17 Ni 12 MO2 and of 2 B finish in accordance with Table 8 of IS:6911

(ii)

Steriliser, Instruments, Table Model IS:5022/1989

Material:

3.1.  All the components other than the drain cock, the element and the thermostat shall be made from stainless steel shall conform to Designation 04Cr18Ni10 of Table 1 and of 2B or 2D finish as given in Table 8 of IS:6911/1972.

 

NOTE:  YOU SHALL BE FULL RESPONSIBLE FOR INSTALLATION, COMMISSIONING AND SATISFACTORY DRAWL FOR AT LEAST 15 CYCLES OVER A PERIOD OF 3 OR 4 DAYS AT CONSIGNEES PREMISES.

Sl.

No.

Item No.

Description of item & Specification

Unit

Rate per unit

(1)

(2)

(3)

(4)

(5)

1.

II(a)

Horizontal Cylindrical Steam Sterilizers Pressure Type              for Hospital and Pharmaceutical use :

 

Pressure type steam sterilizers of horizontal cylindrical type designed for the working pressure of 126 KN/m2  (1.26 Kg/cm2 approx.) but set to operate at 105      (+) 15 KN/m2 (1.05 + 0.15 Kg/cm2 approx.) and the temperature inside is about 121’C with built in steam generator which shall be electrically heated by immersion  heaters having mineral filled sheathed heating elements conforming to IS:4159/2002 duly ISI marked and complying with safety requirements as per IS: 302-2-201 (1992).  The immersion heaters shall be wired for operation in 3 Phase, 4 Wire 400/440 Volts, 50Hz, A/C supply.  The horizontal cylindrical steam sterilizer shall conform to IS:3829(Part 1)/ 1999 with provision of silicon door gasket (expansion type) for the sterilizers and of  the following sizes and with corresponding electric loads :

 

 

Sl.No.

Dia.(mm)

Length(mm)

Electric Load(KW)

i.

400

1100

9

ii.

500

900

9

iii.

500

1100

9

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No.

77,500/-

No.

77,500/-